This section is only relevant to the Progress 3008 (and possibly 3012) cameras.
This should be done every week. The process can be touchy but for
our setup the following seems to work.
On the main screen click the Reset button. This resets the input/output
balances to the factory settings. This should be done at the beginning of
each session at minimum, and may be necessary between individual image-series.
Set the Contrast and Balance sliders to 0 (middle); Set Gain to 0; Exposure
to 1.
Click "Camera". This takes you to the calibration page.
Configure the checkerboard for calibration
Tape the plastic calibration checkerboard to the stage
Position 4 arms WITH CONDENSERS all the way evenly around 2-4 inches
above stage
Aim the 4 arms ("spotlights") at center of grid
Turn power of light sources to around 1/2
Set the magnification to 10-12x
Open diaphragm all the way (5)
Get the checkerboard perfectly in focus
Unfocus the checkerboard slightly (counterclockwise turn of focus)
(this doesn't seem to affect calibration too much)
Setup the calibration
The goal is to get the X(um) and Y(um) values as small and as close together as possible.
Turn the fast scan OFF
Set the integration time to x1 (x1 = exposure 1; x2 = exposure 2;
etc.).
You must calibrate each integration time separately. If
you will not use exposures beyond 1, which is almost always the case, you
don't need calibrate x2, x3 etc. Also if you will not use on or more of
the individual image sizes (768/580; 1024/774; 1928/1450) you need not calibrate
them. In practice you should really only have to calibrate one settings:
Integration time set to x1; Image format 1928/1450 (these settings will provide
the highest quality images).
Click "All" and ensure that all checkboxes are selected
Set the threshold.
The lower the threshold the better
Manufacturer suggests 0.3
We suggest 0.11, lower is possible but difficult to achieve
in the setup at the Chalcidoidea lab
Set the number of iterations.
It should not take more than 20 iterations to get the threshold
to 0.11 low. If it does, the lighting arms position needs to be adjusted
and the power of the light sources may need to be increased or decreased.
If when calibrating either the x or y is consistently much higher than
the other then the camera will not calibrate successfully this time- abort
the calibration and reset the lighting.
Click on the "Start Calibration" button.
Click "Ok" on when asked to setup the slightly focused checkerboard
(done earlier)
Wait for calibration to complete
Hints:
When the lighting is set up properly the camera will calibrate
very quickly, you should know in 4-5 iterations if a successful calibration
will be achieved. You can determined this by watching the change in the
x/y values while the calibration is running. If the values "hover" near your
threshold they will reach it. If your threshold is at 0.1 and the values
consistently are around 0.3 you will likely not calibrate successfully this
round, adjust the lighting and retry (remember to use the condensers!!)
Sometimes
it is difficult to calibrate the Prescan. It should be easily done with
the correct light setting but if it will consistently not calibrate then
deselect and calibrate the rest of the
The camera should now be calibrated.
Double check the "Configure" dialog off of the main dialog:
The matrix should be Ps3200k.mtx for halogen light sources